Series TFXP Transit Time Ultrasonic Flow Meter Operations & Maintenance Manual

Pagina 2

Rev. 5/09 -1.8- TFXP PART 1 - INTRODUCTION The TFXP employs modular construction and provides electr

Pagina 3 - BEFORE OPERATING THE TFXP

Rev. 5/09 -A.4- TFXP UNIT list. Select the appropriate RATE INTERVAL from the list (seconds, minute

Pagina 4 - TABLE OF CONTENTS

Rev. 5/09 -A.5- TFXP Accuracy The accuracy of BTU measurements relies on two separate measurements

Pagina 5

Rev. 5/09 -A.6- TFXP NOTES HEAT FLOW ADDENDUM

Pagina 7

BSC MENU DATALOG LINER TYPE * EBONITE MORTAR HDPE LDPE POLYPRO POLYSTY RUBBER TAR EPXY TEFLON OTHER RATE INT

Pagina 8

CH PSWD? 0-9999 DSP MENU Keypad Operations 1. The (soft)MENU key is pressed from RUN mode to enter PROGRAM mode. The (soft)EXIT key is pressed

Pagina 9

TFX Error Codes Revised 2-22-2002 Code Number Description Correction Warnings 0001 Serial number not present Hardware serial number has become i

Pagina 10 - TFXP -    A -  

K-Factors Explained The K-factor (with regards to flow) is the number of pulses that must be accumulated to equal a particular volume of fluid. You

Pagina 11 - PART 1 - SPECIFICATIONS

Example 2 Known values are: Full Scale Flow Rate = 85 GPM Full Scale Output Frequency = 650 Hz 650 Hz x 60 sec = 39,000 pulses per min 3

Pagina 12 - Installation

4) What is the flow rate in GPM at 4.3 ft/sec. 0.384 gallons x 4.3 FPS x 60 sec (1min) = 99.1 GPM Now that the volumetric flow rate is know all t

Pagina 13 - Connections

Rev. 5/09 -1.9- TFXP PART 1 - SPECIFICATIONS TRANSMITTER Power Requirements Internal 12 V lead-aci

Pagina 14 - Sources

Fluid PropertiesOriginal Date: 7/30/1999Revision: ARevision Date: 9/10/2003File: I:/dynasonics/dyna_code/tables/fluid_ss.xlsFluid Specific Gravity

Pagina 15 - ISO-MODs

Oil, Castor 0.97 1477 4845.8 3.6 0.670 0.649Oil, Diesel 0.80 1250 4101Oil (Lubricating X200) 1530 5019.9Oil (Olive) 0.91 1431 4694.9 2.75 100.000 91.2

Pagina 16 - 4-20 mA

TFX Communications Using UltraLinkThe current implementa tion of the UltraLink software utility was originally developed for Windows95, Windows 98, an

Pagina 17 - ISO-Mods

This converter is required regardless of the operating system or the type of port on the computer.The next step up is the infrared connections used on

Pagina 18

USB (Universal Serial Buss) PortsUSB ports also come in a variety of physical configurations. The two most common are USB Aand USB B. In most computer

Pagina 19

The device manager window should look something like this.Communicating with UltraLinkOnce the com port properties are known the nextstep is to config

Pagina 20 - 1. Mounting Location

The figure on the left shows a typical communications setup for a TFXL, HTTF, or PTTF.The TFXL, HTTF, and PTTF all use astandard RS232 com port type.T

Pagina 21 - Table 2.1

Verifying CommunicationsThere are several possible combinations of indicators that signal communications status. In thelower right hand corner of the

Pagina 22 - 2. Transducer Spacing

Digital Communications Protocol for TFX Flow MetersHost protocolA digital communications protocol is utilized. Each message is guarded with the standa

Pagina 23 - Mounting Mode

Special codesThe following special 65 and 66 codes are supported.Code Description00 Echo (for comm debugging)0A Read run time data (signal strength, f

Pagina 24 - Entering Pipe

Rev. 5/09 -1.10- TFXP PART 1 - TFXP TRANSMITTER CONNECTIONS After unpacking, it is recommended to sav

Pagina 25

C Source CodeFlow Data Definitionstruct FLOWDATA{short sSignalStrength;double dCurFlowRate;double dNetTotalizer;double dPositiveTotalizer;double dNega

Pagina 26 - 3. Transducer Mounting

0x4400, 0x84C1, 0x8581, 0x4540, 0x8701, 0x47C0, 0x4680, 0x8641, 0x8201, 0x42C0, 0x4380, 0x8341, 0x4100, 0x81C1, 0x8081, 0x4040,};unsigned short c

Pagina 27

“STEEL, STAINLESS STEEL, P.V.C. PIPE”STANDARD CLASSESNominal Pipe Size InchesOutside DiameterSCH 5SCH 10(Lt Wall)SCH 20 SCH 30 STD SCH 40ID Wall ID Wa

Pagina 28 - Positioning

“STEEL, STAINLESS STEEL, P.V.C. PIPE”STANDARD CLASSESNominal Pipe Size InchesOutside DiameterSCH 60 X STG. SCH 80 SCH 100 SCH 120/140 SCH 180ID Wall I

Pagina 29

Ductile Iron Pipe (Standard Classes)Size(Inches)ClassMortarLiningSize(Inches)ClassMortarLining50 51 52 53 54 55 56 50 51 52 53 54 55 563”O.D. 3.96 3.9

Pagina 30

Nominal DiameterCOPPER TUBINGCopper & Brass PipeALUMINUMNominal DiameterCOPPER TUBINGCopper & Brass PipeALUMINUMType TypeKLM KLM½”O. D. 0.625

Pagina 31 - Z-Mount

Cast Iron Pipe (Standard Classes)Size(Inches)ClassSize(Inches)ClassABCDEFGH ABCDEFGH3”O.D. 3.80 3.96 3.96 3.9624”O.D.25.80 25.80 26.32 26.32 26.90 26.

Pagina 32

NominalPipe(Inches)I.D. INCH1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5 91 1.05 2.6989 4.0484 5.3978 6.7473 8.097 9.4462 10.796 12.145 13.490 14.8

Pagina 33

NominalPipe(Inches)I.D. INCH1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5 9FPS TO GPM CROSS - REFERENCE (Schedule 40)18 16.88 697.52 1046.3 1395.0 1

Pagina 34 - Mounting

Limited Warranty and Disclaimer Dynasonics, division of Racine Federated Inc. warrants to the end purchaser, for a period of one year from the dat

Pagina 35

Rev. 5/09 -1.11- TFXP 1. The connectors located on the side of the TFXP consist of three 1/4-turn BN

Pagina 36 - 4. RTD INSTALLATION

RETURN OF EQUIPMENT/SALES INFORMATION CONTACTS AND PROCEDURES Customer Service/Application Engineer: If you have a question regarding order status,

Pagina 37 - Insertion RTD

GENERAL TERMS AND CONDITIONS OF SALES 1. PAYMENT – Terms of payment are effective from the actual date of invoice. If, in the Seller’s opinion,

Pagina 38

© 2009 Racine Federated Inc. All rights reserved. Printed in the USA TFXP O&M REV 5/09 DYNASONICS is a registered trademark of Racine Fed

Pagina 39 - Startup

Rev. 5/09 -1.12- TFXP Battery Charging and External Power Sources The 12 volt DC power converter and

Pagina 40 - Operation

Rev. 5/09 -1.13- TFXP NOTE: The TFXP will automatically enter a low power consumption mode approxima

Pagina 41 - Configuration

Rev. 5/09 -1.14- TFXP The standard 4-20 mA output may be replaced with one of the following five ISO-

Pagina 42 - Structure

Rev. 5/09 -1.15- TFXP A 200,000-point Data Logger* is located within the weather-tight pocket on the

Pagina 43 - BSC MENU -- BASIC MENU

Rev. 5/09 -1.16- TFXP PART 1 - INTRODUCTION NOTES

Pagina 44

Rev. 5/09 - 2.1 - TFXP The transducers that are utilized by the TFXP contain piezoelectric c

Pagina 46 - Material

Rev. 5/09 - 2.2 - TFXP The first step in the installation process is the selection of an opt

Pagina 47

Rev. 5/09 - 2.3 - TFXP Table 2.11 — Piping Configuration and Transducer Positioning 1 The

Pagina 48

Rev. 5/09 - 2.4 - TFXP TFX transit time flow meters are sold with four different transducer

Pagina 49 - Heat Capacity

Rev. 5/09 - 2.5 - TFXP Table 2.2 — Transducer Mounting Modes - DTTN / DTTH Transducer Moun

Pagina 50

Rev. 5/09 - 2.6 - TFXP The TFX system calculates proper transducer spacing by utilizing pipi

Pagina 51 - Engineering

Rev. 5/09 - 2.7 - TFXP The following information is required before programming the instrume

Pagina 52

Rev. 5/09 - 2.8 - TFXP After selecting an optimal mounting location (Step 1) and successfull

Pagina 53

Rev. 5/09 - 2.9 - TFXP The spacing between the transducers is measured between the two spaci

Pagina 54

Rev. 5/09 - 2.10 - TFXP 2. Place the upstream transducer in position and secure with a mounti

Pagina 55 - (Optional)

Rev. 5/09 - 2.11 - TFXP 4. If after adjustment of the transducers the signal strength does n

Pagina 56 - 2. DATALOG OPERATION MENU

Rev. 5/09 -1.1- TFXP BEFORE OPERATING THE TFXP The TFXP flow meter is equipped with a lead-acid Gel

Pagina 57 - Duration

Rev. 5/09 - 2.12 - TFXP DTTH High Temperature Transducers Mounting of high temperature trans

Pagina 58

Rev. 5/09 - 2.13 - TFXP 2. On horizontal pipes, mount the transducer in an orientation so th

Pagina 59 - 3. DATALOG MAINTENANCE

Rev. 5/09 - 2.14 - TFXP 3. Crease the paper at the fold line. Mark the crease. Place a mark

Pagina 60

Rev. 5/09 - 2.15 - TFXP 5. For DTTN transducers, place a single bead of couplant, approximate

Pagina 61

Rev. 5/09 - 2.16 - TFXP D010-2102-010 Mounting Track Installation 1. The D010-2102-010 tran

Pagina 62 - 4. OUTPUT 2 MENU

Rev. 5/09 - 2.17 - TFXP 6. Secure with the thumb screw. Ensure that the screw rests in the

Pagina 63

Rev. 5/09 - 2.18 - TFXP For typical installations, the length of RTD wire should equal the le

Pagina 64 - 4-20 mA Span

Rev. 5/09 - 2.19 - TFXP Insertion RTD Installation Insertion RTDs are typically installed th

Pagina 65 - Calibration

Rev. 5/09 - 2.20 - TFXP Figure 2.14 — Insertion RTD Installation Figure 2.15 — Insertion RTD

Pagina 66

Rev. 5/09 -3.1- TFXP NOTE: The TFX flow meter system requires a full pipe of liquid before a successf

Pagina 67 - Rate Pulse

Rev. 5/09 -1.2- TFXP TABLE OF CONTENTS Pages Quick-Start Operating Instructions 1.4 - 1.6 Intro

Pagina 68 - Programming

Rev. 5/09 -3.2- TFXP After installation of the transducers and connection of appropriate cabling to th

Pagina 69 - Dual Relay

Rev. 5/09 -3.3- TFXP “Soft Key” menu items will be displayed immediately above the two keys located in

Pagina 70 - Batch/Totalize

Rev. 5/09 -3.4- TFXP Display Contrast In the RUN mode pressing the SELECT soft key highlights one of t

Pagina 71

Rev. 5/09 -3.5- TFXP Units Selection 5. Sensor Menu – The Sensor menu is used to select the sensor ty

Pagina 72

Rev. 5/09 -3.6- TFXP Transducer Mounting Method XDCR MNT -- Transducer Mounting Method (Choice) 1. V 2

Pagina 73

Rev. 5/09 -3.7- TFXP 16. St Steel 302/303 (SS 303) 17. St Steel 304/316 (SS 316) 18. St Steel 410 (S

Pagina 74 - SEC MENU -- SECURITY MENU

Rev. 5/09 -3.8- TFXP If a pipe material was chosen from the Pipe Mat list, a nominal value for relativ

Pagina 75 - SER MENU -- SERVICE MENU

Rev. 5/09 -3.9- TFXP Liner Sound Speed LINER SS -- Speed of Sound in the Liner (Value) 1. ENGLSH

Pagina 76

Rev. 5/09 -3.10- TFXP FLUID SS -- Speed of Sound in the Fluid (Value) 1. ENGLSH (Feet per Second)

Pagina 77 - DSP MENU -- DISPLAY MENU

Rev. 5/09 -3.11- TFXP If a fluid was chosen from the Fl Type list, a nominal value for specific gravit

Pagina 78 - Timeout

Rev. 5/09 -1.3- TFXP TABLE OF CONTENTS Pages Keypad Entry Detail 3.4 - 3.33 BASIC MENU 3.4 -

Pagina 79 - PART 4 - SOFTWARE UTILITIES

Rev. 5/09 -3.12- TFXP Table 3.2 — Specific Heat Capacity Values for Ethylene Glycol/Water Specific He

Pagina 80

Rev. 5/09 -3.13- TFXP Transducer Spacing XDCR SPAC -- Transducer Spacing Calculation (Value) 1. ENGLS

Pagina 81

Rev. 5/09 -3.14- TFXP Engineering Units TOTALIZER Engineering Units RATE INTERVAL 8. Liquid Barrels (

Pagina 82

Rev. 5/09 -3.15- TFXP Engineering Units TOTAL Exponent Minimum Flow Rate 13. Millions of BTUs (MMBTU)

Pagina 83

Rev. 5/09 -3.16- TFXP MAX RATE -- Maximum Flow Rate Settings (Value) Rate Unit/Rate Interval A maximu

Pagina 84

Rev. 5/09 -3.17- TFXP I/O module bay number 1 is reserved exclusively for the ISO Data Logger module.

Pagina 85

Rev. 5/09 -3.18- TFXP File Number or Location Data Logger Configuration NOTE: The logger will not aut

Pagina 86

Rev. 5/09 -3.19- TFXP DATALOG INT -- Datalog Interval (Value) ((Sec)) 1 to 30,000 From the Datalog Op

Pagina 87

Rev. 5/09 -3.20- TFXP Also included with the data logging package is a stand-alone power supply for us

Pagina 88

Rev. 5/09 -3.21- TFXP Datalog Maintenance permits files to be deleted from the data logger module. Th

Pagina 89

Rev. 5/09 -1.4- TFXP This manual contains detailed operating instructions for all aspects of the TFX

Pagina 90

Rev. 5/09 -3.22- TFXP This number will appear in the “Name” column on the computer screen before data

Pagina 91

Rev. 5/09 -3.23- TFXP NOTES: Data logger must be stopped before downloading data. If data logger is no

Pagina 92

Rev. 5/09 -3.24- TFXP ISO-MOD: 4-20 mA Module (Optional) The second I/O bay is internal to the TFXP

Pagina 93

Rev. 5/09 -3.25- TFXP The 4-20 mA output can be configured to source current (internally powered) or s

Pagina 94

Rev. 5/09 -3.26- TFXP 4-20 mA Span External Power Configuration (Current Sink): If the 4-20 mA output

Pagina 95

Rev. 5/09 -3.27- TFXP 4-20 mA Calibration Example 1: To span the 4-20 mA output from -100 GPM to +100

Pagina 96

Rev. 5/09 -3.28- TFXP NOTE: The CAL 4MA and CAL 20MA entries should not be used in an attempt to set

Pagina 97

Rev. 5/09 -3.29- TFXP 4-20 mA Test ISO-MOD: Rate Pulse Module (Optional) 4-20 TST -- 4-20mA Output T

Pagina 98

Rev. 5/09 -3.30- TFXP Example 2: To span the 0 to 2.5k Hz output from 0 GPM to +100 GPM, with 1.25k H

Pagina 99

Rev. 5/09 -3.31- TFXP ISO-MOD: Dual Relay Module (Optional) RELAY -- Dual Relay (Choices and Values)

Pagina 100 - HEAT FLOW ADDENDUM

Rev. 5/09 -1.5- TFXP 2. TRANSDUCER/POWER CONNECTIONS A. Route the transducer cables from the transdu

Pagina 101

Rev. 5/09 -3.32- TFXP Flow Rate Relay Batch/Totalize Mode Totalize mode configures the relay(s) to out

Pagina 102

Rev. 5/09 -3.33- TFXP Signal Strength Alarm Errors Alarm Relay The Sig Str alarm will provide an indi

Pagina 103 - APPENDIX

Rev. 5/09 -3.34- TFXP ISO-MOD: RS232C Module (Optional) RS232C -- RS232C (Choice) Baud Rate (RS23

Pagina 104 - DATALOG

Rev. 5/09 -3.35- TFXP ISO-MOD: RS485 Module (Optional) RS485 -- RS485 (Choices and Values) RS485 Mode

Pagina 105 - DSP MENU

Rev. 5/09 -3.36- TFXP System RESET Totalizer RESET The SEN MENU is utilized to select the type of tran

Pagina 106 - TFX Error Codes

Rev. 5/09 -3.37- TFXP Substitute Flow Entry Change Password Signal Strength Cutoff CH PSWD? -- Change

Pagina 107 - 42,000 pulses per min

Rev. 5/09 -3.38- TFXP Setting/Calibrating Zero Flow SET ZERO -- Calibrating Zero Flow (Choice) 1. No

Pagina 108 - 39,000 pulses per min

Rev. 5/09 -3.39- TFXP Correction Factor Graphics Display Mode the meter. This is not as desirable as t

Pagina 109 - = 99.1 GPM

Rev. 5/09 -3.40- TFXP Back Light Timeout BACK LIGHT -- Back Light Timeout (Choice) 1. 30 Seconds (

Pagina 110 - Fluid Properties

Rev. 5/09 -4.1- TFXP PART 4 - SOFTWARE UTILITIES The TFXP flow meter can be used with two software utilities, ULT

Pagina 111

Rev. 5/09 -1.6- TFXP Startup QUICK-START OPERATING INSTRUCTIONS C. Apply a single 1/2" (12 mm)

Pagina 112

Rev. 5/09 -4.2- TFXP Initialization Connect the infrared communications adapter to a PC communication port and p

Pagina 113

Rev. 5/09 -4.3- TFXP The first screen that appears after clicking the Configuration button is the BASIC tab. See

Pagina 114

Rev. 5/09 -4.4- TFXP Transducer Mount selects the orientation of the transducers on the piping system. See Part

Pagina 115

Rev. 5/09 -4.5- TFXP Flow Tab Flow Rate Units are selected from the pull-down lists. Select an appropriate rate

Pagina 116

Rev. 5/09 -4.6- TFXP The Damping value is increased to improve stability of the flow rate readings. Damping valu

Pagina 117

Rev. 5/09 -4.7- TFXP Substitute Flow is used to provide an indication and output that signifies that an error exi

Pagina 118 - Host protocol

Rev. 5/09 -4.8- TFXP Time Domain Filter adjusts the number of raw data sets (the wave forms viewed on the ULTRALI

Pagina 119

Rev. 5/09 -4.9- TFXP Flow Filter MinHysteresis sets a minimum hysteresis window that is invoked at sub 0.25 FPS (

Pagina 120

Rev. 5/09 -4.10- TFXP Output Tab The entries made in the Output tab establish input and output calibration and r

Pagina 121

Rev. 5/09 -4.11- TFXP Calibration/Test is used to adjust the factory calibration span of the 4-20 mA output and t

Pagina 122 - STANDARD CLASSES

Rev. 5/09 -1.7- TFXP PART 1 - INTRODUCTION The TFXP ultrasonic flow meter is designed to measure the

Pagina 123

Rev. 5/09 -4.12- TFXP Setting Zero and Calibration ULTRALINK™ contains a powerful multi-point calibration routi

Pagina 124

Rev. 5/09 -4.13- TFXP The final screen (Page 3 of 3) shown in Figure 4.9 on page 4.14 allows multiple actual flow

Pagina 125 - ALUMINUM

Rev. 5/09 -4.14- TFXP Saving the Configuration Printing a Report Saving Meter Configuration on a PC The complete

Pagina 126

Rev. 5/09 -4.15- TFXP During the installation of ULTRALINK™, a file called Data Logger was installed and its icon

Pagina 127

Rev. 5/09 -4.16- TFXP Figure 4.11 — File Uploading Figure 4.12 —

Pagina 128

Rev. 5/09 -4.17- TFXP The data logger module contains a real-time clock that can be set by clicking the Clock but

Pagina 129

Rev. 5/09 -4.18- TFXP NOTES PART 4 - SOFTWARE UTILITIES

Pagina 130 - CONTACTS AND PROCEDURES

Rev. 5/09 -A.1- TFXP The TFX flow meter with the optional RTD (heat flow) Module installed is design

Pagina 131

Rev. 5/09 -A.2- TFXP Installation Installation 1. Follow the instructions outlined in the standard

Pagina 132

Rev. 5/09 -A.3- TFXP 5. Route a cable from the electrical junction box back to the TFX flow meter.